Details
Used for the quantitative determination of NiV Fusion protein concentration in Human/Mouse serum, plasma and cell culture supernates.
This assay employs the quantitative sandwich enzyme immunoassay technique. An antibody specific for NiV Fusion protein has been pre-coated onto a microplate. Standards or samples are pipetted into the wells and any NiV Fusion protein present is bound by the immobilized antibody. After washing away any unbound substances, a biotin-labeled antibody specific for NiV Fusion protein is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of NiV Fusion protein bound in the initial step. The color development is stopped and the intensity of the color is measured.
Intra-Assay Precision (Precision within an assay): <10% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <15% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
| Sample | 1 | 2 | 3 | 1 | 2 | 3 | |
| n | 16 | 16 | 16 | 24 | 24 | 24 | |
| Mean (ng/mL) | 10.5 | 2.2 | 0.7 | 10.5 | 2.1 | 0.7 | |
| Standard deviation | 0.6 | 0.1 | 0.1 | 0.7 | 0.1 | 0.1 | |
| CV \(%\) | 5.8 | 4.9 | 9.8 | 6.2 | 4.5 | 12.9 |
The recovery of NiV Fusion protein spiked to three different levels in human serum samples, mouse serum samples and cell culture supernates throughout the range of the assay was evaluated.
| Sample | Dilution factor | Average (%) | Range (%) |
| Human serum (n=3) | 1:10 | 87 | 81-94 |
| 1:20 | 90 | 82-109 | |
| 1:10 | 90 | 83-105 | |
| 1:20 | 99 | 90-105 | |
| 1:2 | 92 | 85-97 |
To assess the linearity of the assay, human serum samples, mouse serum samples and cell culture supernates were spiked with high concentrations of NiV Fusion protein and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
| Sample | Dilution factor | Average (%) | Range (%) |
| Human serum (n=3) | 1:2 | 111 | 103-117 |
| 1:4 | 117 | 113-120 | |
| 1:8 | 103 | 100-107 | |
| 1:16 | 100 | 85-112 | |
| 1:2 | 107 | 101-110 | |
| 1:4 | 107 | 101-113 | |
| 1:8 | 98 | 82-113 | |
| 1:16 | 105 | 95-116 | |
| 1:2 | 103 | 101-105 | |
| 1:4 | 107 | 106-108 | |
| 1:8 | 110 | 107-113 | |
| 1:16 | 105 | 97-113 |