Sunlong Medical™ Pseudomonas fluorescens Host Cell Protein (HCP) G3 Residual Detection ELISA Kit (One-Step)
For research use only. Not for use in diagnostic procedures.
1. APPLICATION
For the quantitative determination of host cell protein (HCP) residues of Pseudomonas fluorescens in culture supernatant, protein purification processes and final products. The kit is suitable for in-process and final-product samples and is intended for the monitoring of host cell protein clearance during downstream process development and quality control.
This kit is intended for the quantitative analysis of residual host cell protein (HCP) in protein drugs when Pseudomonas fluorescens is used as the expression system.
2. PRINCIPLE
This kit is based on the solid-phase sandwich enzyme-linked immunosorbent assay (ELISA). A capture antibody is pre-coated onto the microplate to form the solid-phase antibody. During the assay, the antigen to be tested and a horseradish peroxidase (HRP)-labeled detection antibody are added simultaneously to the coated wells to form a capture antibody-antigen-detection antibody complex. After washing, unbound conjugates are removed, and the substrate TMB is added for color development. TMB is oxidized by HRP and turns blue, and then turns yellow in the presence of acid. The color intensity after oxidation is positively correlated with the total amount of the factor. The absorbance (OD value) is measured on a microplate reader at dual wavelengths of 450 nm and 630 nm, fitted against the concentration to construct a standard curve, and the sample concentration is calculated by substituting the sample OD value into the standard curve equation.
3. KIT SPECIFICATIONS
Sensitivity (LOD): 0.39 ng/mL
Limit of quantification (LOQ): 1.56 ng/mL
Assay range: 1.56 - 100 ng/mL (7-point standard curve, S1 = 100 ng/mL)
Precision: intra-assay CV < 5%; inter-assay CV < 6%
Sample volume: 50 μL per well
Assay type: solid-phase sandwich ELISA (one-step), read at 450 nm
Specificity: the antibodies recognise HCP with a coverage of more than 80% and the antiserum titer of the antibody used in the kit reaches 10⁶ as determined by indirect ELISA; the antibody is purified by affinity purification to remove non-specific antibodies to the greatest extent
Recovery: sample-matrix dependent; a spike recovery test is recommended when a different buffer is used for the first time, in order to exclude matrix interference effects; acceptable spike recovery 80 - 120%; in practice, mix the diluted standard S1 (100 ng/mL) with the sample matrix at a 1:1 volume ratio
4. MATERIALS PROVIDED
For items that differ between the two kit sizes, values are given as 48 Test / 96 Test.
| # | Component | Specification |
| 1 | Antibody-coated microplate (detachable) | 48 wells / 96 wells |
| 2 | Standard stock solution (500 μg/mL) | 30 μL / 50 μL |
| 3 | Detection Antibody stock (100×) | 100 μL / 150 μL |
| 4 | TMB Substrate | 10 mL |
| 5 | Stop Solution | 10 mL |
| 6 | Wash Buffer Concentrate (20×) | 10 mL |
| 7 | Diluent (10×) | 10 mL |
| 8 | Plate Cover Seals | - |
| 9 | Instructions | - |
5. ASSAY PROCEDURE SUMMARY
Step 1. Add 50 μL of each standard or sample and 100 μL of 1× Detection Antibody to the appropriate wells; seal the plate and incubate at room temperature with shaking for 2 hours.
Step 2. Wash the plate 5 times with 1× Wash Buffer.
Step 3. Add 100 μL of TMB Substrate Solution and incubate for about 15 minutes at room temperature in the dark.
Step 4. Add 50 μL of Stop Solution and read the OD value at 450 nm.
6. TYPICAL DATA
Appendix I: Example of a standard curve (schematic; for reference only).