Details
Immunoglobulins (Ig), also known as antibodies, are glycoproteins produced by B cells and plasma cells. IgG is the most common type of antibody found in blood circulation. It is composed of two identical light chains and two identical heavy chains that form a disulfide-linked Y-shaped molecule. IgG consists of four subclasses in mice (IgG1, IgG2a/2c, IgG2b, and IgG3) and plays a crucial role in opsonizing pathogens for phagocytosis and activating the complement system. IgG2a and IgG2c were reported as the allelic variants in most mouse strains. IgG2c was observed in the C57BL/6 mouse serum, while the BALB/c mouse serum had IgG2a but not IgG2c.
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
| Sample | n | Mean (ng/mL) | SD | CV% |
| 1 | 8 | 5.51 | 0.17 | 3.09 |
| 2 | 8 | 1.29 | 0.06 | 4.65 |
| 3 | 8 | 0.71 | 0.02 | 2.82 |
Inter-assay Precision
| Sample | n | Mean (ng/mL) | SD | CV% |
| 1 | 16 | 5.50 | 0.15 | 2.73 |
| 2 | 16 | 1.28 | 0.05 | 3.91 |
| 3 | 16 | 0.70 | 0.02 | 2.86 |
The recovery of mouse IgG2c spiked to three different levels throughout the range of the assay in C57BL/6 mouse serum was evaluated.
| Sample Type | Average% of Expected | Range (%) | |
| C57BL/6 mouse serum | 1:320,000 | 100 | 94-105 |
| 1:640,000 | 108 | 103-115 |
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The C57BL/6 mouse serum was initially diluted 1:40,000.)
| C57BL/6 mouse serum | ||
| 1:2 | Average% of Expected | 100 |
| Range (%) | - | |
| 1:4 | Average% of Expected | 97 |
| Range (%) | 96-98 | |
| 1:8 | Average% of Expected | 92 |
| Range (%) | 89-95 | |
| 1:16 | Average% of Expected | 90 |
| Range (%) | 86-94 |
This immunoassay is calibrated against a highly purified recombinant mouse IgG2c produced at Sunlong Medical™.