MAG encodes a glia-specific adhesion and signaling receptor that anchors the axon-myelin interface, maintaining axonal integrity and limiting aberrant regeneration. Localized to the periaxonal membrane of myelinating glia, MAG regulates myelin stability, axon survival, and growth inhibition via Siglec-dependent pathways. Dysregulation or autoantibody targeting of MAG contributes to demyelinating neuropathies, neurodegeneration, and impaired axonal repair.
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
| Sample | n | Mean (ng/mL) | SD | CV% |
| 1 | 8 | 17.19 | 0.19 | 1.11 |
| 2 | 8 | 3.91 | 0.05 | 1.28 |
| 3 | 8 | 2.06 | 0.06 | 2.91 |
Inter-assay Precision
| Sample | n | Mean (ng/mL) | SD | CV% |
| 1 | 16 | 17.11 | 0.34 | 1.99 |
| 2 | 16 | 3.97 | 0.08 | 2.02 |
| 3 | 16 | 2.05 | 0.06 | 2.93 |
The recovery of mouse MAG spiked to three different levels throughout the range of the assay in tissue lysate was evaluated.
| Sample Type | Average% of Expected | Range (%) | |
| Tissue lysate | 1:200 | 95 | 91-97 |
| 1:400 | 102 | 98-104 |
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The tissue lysate was initially diluted 1:100.)
| Tissue lysate | ||
| 1:2 | Average% of Expected | 100 |
| Range (%) | - | |
| 1:4 | Average% of Expected | 104 |
| Range (%) | 94-115 | |
| 1:8 | Average% of Expected | 95 |
| Range (%) | 86-104 | |
| 1:16 | Average% of Expected | 104 |
| Range (%) | 91-117 |
This immunoassay is calibrated against a highly purified recombinant human Mouse MAG One-Step ELISA Kit produced at Sunlong Medical™.