Details
IgM is one of several isotypes of antibody (immunoglobulin) produced by the immune system and is the first antibody to be generated in response to an initial exposure to an antigen. In clinical medicine, the detection of IgM antibodies against a specific pathogen (such as a virus or bacteria) usually indicates a recent infection or the acute phase of a disease. As the immune response progresses, IgM levels typically decline and are replaced by longer-lasting IgG antibodies.
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
| Sample | n | Mean (ng/mL) | SD | CV% |
| 1 | 8 | 5.21 | 0.24 | 4.61 |
| 2 | 8 | 1.30 | 0.03 | 2.31 |
| 3 | 8 | 0.65 | 0.05 | 7.69 |
Inter-assay Precision
| Sample | n | Mean (ng/mL) | SD | CV% |
| 1 | 16 | 5.55 | 0.62 | 11.17 |
| 2 | 16 | 1.35 | 0.08 | 5.93 |
| 3 | 16 | 0.66 | 0.05 | 7.58 |
The recovery of mouse IgM spiked to three different levels throughout the range of the assay in mouse serum was evaluated.
| Sample Type | Average% of Expected | Range (%) | |
| Mouse serum | 1:80,000 | 97 | 89-105 |
| 1:160,000 | 102 | 99-105 |
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The mouse serum was initially diluted 1:20,000.)
| Mouse serum | ||
| 1:2 | Average% of Expected | 100 |
| Range (%) | - | |
| 1:4 | Average% of Expected | 88 |
| Range (%) | 84-92 | |
| 1:8 | Average% of Expected | 89 |
| Range (%) | 88-91 | |
| 1:16 | Average% of Expected | 100 |
| Range (%) | 89-110 |