Details
Immunoglobulin A (IgA) is a crucial antibody that plays a pivotal role in the immune system's first line of defense, primarily safeguarding the body's mucosal surfaces-such as the respiratory, gastrointestinal, and urogenital tracts. As the second most abundant isotype in human serum and the dominant antibody found in external secretions like saliva, tears, and breast milk, IgA specialized in neutralizing pathogens, preventing toxins and microbes from adhering to epithelial cells, and maintaining a healthy balance with the body's natural microbiota.
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
| Sample | n | Mean (ng/mL) | SD | CV% |
| 1 | 8 | 9.39 | 0.48 | 5.11 |
| 2 | 8 | 2.58 | 0.19 | 7.36 |
| 3 | 8 | 1.36 | 0.02 | 1.47 |
Inter-assay Precision
| Sample | n | Mean (ng/mL) | SD | CV% |
| 1 | 16 | 9.39 | 0.40 | 4.26 |
| 2 | 16 | 2.58 | 0.14 | 5.43 |
| 3 | 16 | 1.39 | 0.04 | 2.88 |
The recovery of human IgA spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average% of Expected | Range (%) | |
| Human plasma | 1:1,600,000 | 103 | 101-106 |
| 1:3,200,000 | 110 | 109-111 | |
| Urine | 1:320 | 93 | 89-98 |
| 1:640 | 97 | 96-98 | |
| Saliva | 1:32,000 | 91 | 88-97 |
| 1:64,000 | 103 | 94-117 | |
| Human milk | 1:80,000 | 95 | 92-99 |
| 1:160,000 | 99 | 97-101 |
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human plasma was initially diluted 1:200,000. The urine was initially diluted 1:40. The saliva was initially diluted 1:4,000. The human milk was initially diluted 1:20,000.)
| Human plasma | Urine | Saliva | Human milk | ||
| 1:2 | Average% of Expected | 100 | 100 | 100 | 100 |
| Range (%) | - | - | - | - | |
| 1:4 | Average% of Expected | 103 | 102 | 98 | 100 |
| Range (%) | 101-105 | 101-103 | 94-101 | 100-101 | |
| 1:8 | Average% of Expected | 103 | 98 | 101 | 103 |
| Range (%) | 101-104 | 97-99 | 98-104 | 100-105 | |
| 1:16 | Average% of Expected | 109 | 98 | 106 | 102 |
| Range (%) | 104-113 | 95-101 | 102-110 | 100-104 |
This immunoassay is calibrated against a highly purified recombinant human IgA produced at Sunlong Medical™.