Details
Connective tissue growth factor (CTGF) is also named as CCN family member 2 (CCN2) and Cellular communication network factor 2. CTGF regulates critical biological activities like cell growth, invasion, and migration, impacting breast cancer development and progression. It modulates breast tumor microenvironment by promoting angiogenesis, activating cancer-associated fibroblasts (CAFs), and inducing inflammation. CTGF can regulate breast cancer cells by activating various signaling pathways and modulating the transcription of other genes that are involved in tumor development and metastasis including S100A4, glucose transporter 3 (GLUT3), and vascular endothelial growth factor (VEGF). CTGF exerts a broad impact on cell migration, adhesion and proliferation, establishing it as a critical element in the development of fibrotic diseases such as renal fibrosis, cardiac fibrosis, and pulmonary fibrosis, among others.
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
| Sample | n | Mean (pg/mL) | SD | CV% |
| 1 | 8 | 2,438.0 | 69.1 | 2.8 |
| 2 | 8 | 587.6 | 24.3 | 4.1 |
| 3 | 8 | 289.6 | 12.4 | 4.3 |
Inter-assay Precision
| Sample | n | Mean (pg/mL) | SD | CV% |
| 1 | 16 | 2,461.9 | 80.7 | 3.3 |
| 2 | 16 | 597.7 | 24.7 | 4.1 |
| 3 | 16 | 297.7 | 12.8 | 4.3 |
The recovery of rat CTGF spiked to three different levels throughout the range of the assay was evaluated.
| Sample Type | Average% of Expected | Range (%) | |
| Rat serum | 1:320 | 100 | 91-104 |
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The rat serum was initially diluted 1:20. )
| Rat serum | ||
| 1:2 | Average% of Expected | 100 |
| Range (%) | - | |
| 1:4 | Average% of Expected | 86 |
| Range (%) | 84-88 | |
| 1:8 | Average% of Expected | 88 |
| Range (%) | 79-96 | |
| 1:16 | Average% of Expected | 82 |
| Range (%) | 72-92 |
This immunoassay is calibrated against a highly purified recombinant rat CTGF produced at Sunlong Medical™.
Ordering & shipping
Citations
Partial purchase records (0)
Leave a message
+86 0571 56623320
[email protected]
+86 18668110335