Details
hCG Beta (β-hCG) is the beta subunit of human chorionic gonadotropin (hCG), a glycoprotein hormone that belongs to the same family. The hCG heterodimer consists of a common alpha subunit and a hormone-specific beta subunit; the beta subunit confers biological specificity and distinguishes hCG from other glycoprotein hormones. While the alpha subunit is essentially identical across this hormone family, the hCG beta subunit is unique, featuring a carboxy-terminal extension of approximately 30 amino acids with associated carbohydrate residues that distinguish it from the LH beta subunit. Following implantation, hCG is produced primarily by the syncytiotrophoblast of the placenta and maintains the corpus luteum to sustain progesterone production during early pregnancy. In clinical practice, the beta subunit serves as the basis for most modern hCG immunoassays because antibodies directed against it minimize cross-reactivity with LH, enabling sensitive and specific detection of pregnancy and hCG-secreting tumors such as gestational trophoblastic disease and germ cell tumors.
Sandwich ELISA structure (Detection antibody labeled with HRP) A capture antibody is pre-coated onto the bottom of wells which binds to analyte of interest. A detection antibody labeled with HRP also binds to the analyte. TMB acts as the HRP substrate and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns solution yellow. The color intensity is proportional to the quantity of bound protein which is measurable at 450 nm with the correction wavelength set at 630 nm.
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
| Sample | n | Mean (pg/mL) | SD | CV% |
| 1 | 8 | 2,673.0 | 135.4 | 5.1 |
| 2 | 8 | 672.6 | 36.2 | 5.4 |
| 3 | 8 | 374.1 | 15.3 | 4.1 |
Inter-assay Precision
| Sample | n | Mean (pg/mL) | SD | CV% |
| 1 | 16 | 2,632.3 | 124.7 | 4.7 |
| 2 | 16 | 689.7 | 35.0 | 5.1 |
| 3 | 16 | 363.2 | 22.4 | 6.2 |
The recovery of hCG Beta spiked to three different levels throughout the range of the assay in human plasma was evaluated.
| Sample Type | Average% of Expected | Range (%) | |
| Human plasma | 1:128 | 107 | 101-114 |
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human plasma was initially diluted 1:8.)
| Human plasma | ||
| 1:2 | Average% of Expected | 100 |
| Range (%) | - | |
| 1:4 | Average% of Expected | 93 |
| Range (%) | 91-96 | |
| 1:8 | Average% of Expected | 86 |
| Range (%) | 82-89 | |
| 1:16 | Average% of Expected | 82 |
| Range (%) | 79-84 |