Details
Signal regulatory protein Alpha (SIRP Alpha), also known as CD172a, was first identified as a membrane protein present mainly on macrophages and myeloid cells. SIRP Alpha recognizes as ligand CD47, a receptor expressed on all normal cells, but often overexpressed on cancer cells. The interaction plays a role in controlling phagocytosis, and CD47 is often termed a "don't-eatme" signal through its effect on SIRP Alpha.
Sandwich ELISA structure (Detection antibody labeled with HRP) A capture antibody is pre-coated onto the bottom of wells which binds to analyte of interest. A detection antibody labeled with HRP also binds to the analyte. TMB acts as the HRP substrate and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns solution yellow. The color intensity is proportional to the quantity of bound protein which is measurable at 450 nm with the correction wavelength set at 630 nm.
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
| Sample | n | Mean (pg/mL) | SD | CV% |
| 1 | 8 | 2,220.1 | 66.5 | 3.0 |
| 2 | 8 | 506.0 | 10.3 | 2.0 |
| 3 | 8 | 268.2 | 2.1 | 0.8 |
Inter-assay Precision
| Sample | n | Mean (pg/mL) | SD | CV% |
| 1 | 16 | 2,257.6 | 104.7 | 4.6 |
| 2 | 16 | 502.8 | 10.5 | 2.1 |
| 3 | 16 | 271.0 | 3.8 | 1.4 |
The recovery of human SIRP Alpha/CD172a spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average% of Expected | Range (%) | |
| Human serum | 1:80 | 88 | 87-90 |
| 1:160 | 94 | 89-97 | |
| Cell lysate | 1:16 | 103 | 94-114 |
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human serum was initially diluted 1:10. The cell lysate was initially diluted 1:2.)
| Human serum | Cell lysate | ||
| 1:2 | Average% of Expected | 100 | 100 |
| Range (%) | - | - | |
| 1:4 | Average% of Expected | 101 | 87 |
| Range (%) | 96-105 | 85-88 | |
| 1:8 | Average% of Expected | 99 | 79 |
| Range (%) | 89-109 | 79-80 | |
| 1:16 | Average% of Expected | 89 | 78 |
| Range (%) | 81-97 | 74-82 |