Product Summary

NO.:EL0340Hu

Sensitivity:0.78pg/mL

Assay range:3.125pg/mL-100pg/mL

Assay type:Sandwich ELISA

Validity:Six Monthes

Species:Human

    Number:

Details

Assay Principle

This kit employs double antibody sandwich ELISA technology: Capture Antibody is coated on the microplate to capture RLN-3 from samples and standards. After washing, biotin-labeled Detection Antibody is added and incubated, followed by washing to form a "Capture Antibody-Antigen-Detection Antibody" immune-complex. Subsequently, Streptavidin-Horseradish Peroxidase (SA-HRP) is added and incubated. After incubation and washing, TMB Substrate is added for color development. If the target analyte is present in the sample, a blue color develops. Stop Solution is then added to terminate the reaction. During the assay, unbound components are washed away. The Optical Density (OD) is measured at 450 nm using a microplate reader. The intensity of the color is proportional to the RLN-3 concentration in the sample, and the concentration is calculated by plotting a standard curve.

Precision

Intra-assay and inter-assay Coefficients of Variation (CV) are both <10%.

· Intra-assay Precision: Three known concentration samples were assayed 20 times on one plate. The CV of the concentrations was calculated.

· Inter-assay Precision: Three known concentration samples were assayed in 20 replicates across three different plates. The CV of the concentrations was calculated.

 

Intra-assay Precision

Inter-assay Precision

Sample

1

2

3

1

2

3

n

20

20

20

20

20

20

Mean (pg/mL)

3.69

6.44

13.35

4.1

5.71

11.63

Standard Deviation

0.03

0.15

0.42

0.07

0.17

0.4

CV (%)

0.81

2.32

3.14

1.70

2.97

3.43

Recovery

Recovery was tested by spiking known concentrations of Human RLN-3 into different sample matrices. The recovery range and average recovery are shown below.

 

Sample Type

Range (%)

Average Recovery (%)

Serum (n=8)

90-102

96

Plasma (n=8)

92-104

98

Cultured Cells (n=8)

98-115

109