Details
Summary:
Purpose
Our Quick Step Human Claudin 11 (CLDN11)ELISA Kit is to assay CLDN11 levels in Human serum, plasma, culture media or any biological fluid.
Principle
This ELISA kit uses Sandwich-ELISA as the method. The Microelisa stripplate provided in this kit has been pre-coated with an antibody specific to CLDN11. Standards or samples are added to the appropriate Microelisa stripplate wells and combined to the specific antibody. Then a Horseradish Peroxidase (HRP)-conjugated antibody specific for CLDN11 is added to each Microelisa stripplate well and incubated. Free components are washed away. The TMB substrate solution is added to each well. Only those wells that contain CLDN11 and HRP conjugated CLDN11 antibody will appear blue in color and then turn yellow after the addition of the stop solution. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of CLDN11. You can calculate the concentration of CLDN11 in the samples by comparing the OD of the samples to the standard curve.
Typical Data
The standard curve of QS4482Hu is provided for demonstration only. A standard curve should be generated for each set of samples assayed.
Standard |
Concentration |
OD Value |
Average OD Value |
Blank Well |
0pg/ml |
0.049 |
0.0440 |
0.039 |
|||
S1 |
1800pg/ml |
1.956 |
1.9530 |
1.950 |
|||
S2 |
1200pg/ml |
1.389 |
1.3910 |
1.393 |
|||
S3 |
600pg/ml |
0.709 |
0.7135 |
0.718 |
|||
S4 |
300pg/ml |
0.390 |
0.3875 |
0.385 |
|||
S5 |
150pg/ml |
0.195 |
0.2000 |
0.205 |