AK0397-100T-96S Pyruvate Assay Kit Linear Range
2026-11-17
|
Technical Q & A
AK0397-100T-96S Pyruvate Assay Kit Linear Range
|
| Product | Pyruvate (PA) Content Assay Kit (Pyruvate PA Content Assay Kit) |
| Cat. No. | AK0397-100T-96S(规格:100T-96S) |
Using this pyruvate content assay kit, with a 20 µmol/mL sodium pyruvate standard and a 0.125 µmol/mL working standard. Question: Before applying the serum/plasma extraction correction factor, does the measured extracted sample solution at 520 nm apply to the linear range 0.0015–0.25 µmol/mL, or does the original serum/plasma concentration after applying the correction factor apply? The customer uses 100 µL serum/plasma with 1000 µL extraction reagent, corresponding to an 11-fold correction factor.
The pyruvate concentration in the sample supernatant added to the assay tube only needs to be within the linear range of 0.0015-0.25 µmol/mL; that is, the value obtained by ∆Aassay ÷ ∆Astandard × Cstandard only needs to be within 0.0015-0.25.
· Sample type: serum/plasma extracted sample solution (prepared from 100 µL serum/plasma and 1000 µL extraction reagent, corresponding to an 11-fold correction factor)
· Other limitations: The pyruvate concentration in the sample supernatant added to the assay tube only needs to be within the linear range of 0.0015–0.25 µmol/mL, i.e., the value obtained by ∆A_test ÷ ∆A_standard × C_standard is within 0.0015–0.25.